Guide · Implant prosthetics
Implant screw loosening: causes and what to do
Why screws on implant crowns and full-arch bridges come loose, how to tell a loose screw from a debonded crown, and what to send the lab for a repair.
Short answer
Implant screw loosening happens when the screw loses preload, the clamping force that holds the restoration against the implant. The causes most often cited are under-torquing, a component that never fully seated, off-axis loading, cantilevers and bruxism. Retighten with a torque driver to the manufacturer's value, then deal with the cause before it loosens again.
Step by step
- 01
Work out what is actually loose
Before you remove anything, watch what moves: the whole restoration, just the crown on its Ti-base, or the implant itself. A loose screw lets the whole assembly move; a debonded crown spins on a Ti-base that stays tight; a mobile implant is a different workup.
- 02
Take a periapical square to the connection
Aim the beam perpendicular to the implant-abutment interface. A visible gap there means the component never seated, and retightening over it won't fix the problem.
- 03
Remove the restoration and inspect screw and components
Check the screw head and threads, the Ti-base and the connection. Worn threads, a stripped head or a debonded Ti-base change the plan. Bag and label the parts, because the lab will need to see them.
- 04
Retighten with a torque driver to the manufacturer's value
Use the torque in the component manufacturer's instructions for use, always with a torque driver. One in vitro study recommends retightening 10 minutes after the first torque to recover the preload lost as the surfaces settle.
- 05
Adjust the occlusion before the patient leaves
Mark contacts with articulating paper and shim stock. On a posterior implant crown, keep contact to firm closure only and remove lateral and protrusive contacts. Retightening doesn't fix an occlusal cause.
- 06
Protect the screw head and chart the torque
Cover the screw head with PTFE tape before sealing the access so the next re-entry doesn't damage it. Note the torque and date in the chart and check the screw at the next visit.
- 07
If it loosens again, send the whole assembly to the lab
A second loosening means the cause isn't just torque. Talk to the lab and send the restoration with its Ti-base and screw, the radiograph, photos of the marked contacts and the implant brand and platform.
Why do implant screws come loose?
An implant screw comes loose when its preload, the tension created at tightening that clamps the restoration to the implant, drops below the forces trying to pull that joint apart during chewing. Low torque, a restoration that never fully seated and lateral loading all eat into that margin.
It is one of the most common technical complications. In Jung and colleagues’ systematic review (Clinical Oral Implants Research, 2012) of 46 studies on implant single crowns, the cumulative five-year incidence of screw loosening was 8.8%.
Huang and Wang (Journal of Esthetic and Restorative Dentistry, 2019) reviewed 99 papers on the mechanism. Internal connections and abutments with conical, anti-rotational designs resist loosening better; cantilevers raise the risk. Their clinical advice: torque to the recommended value, avoid tightening and loosening the same screw over and over, and shorten recall intervals so loose screws get caught early.
Is it the screw, the crown or the implant?
A loose screw, a debonded crown and a mobile implant all produce the same complaint, “my crown moves”, and each needs a different response. In Stilwell’s review (British Dental Journal, 2024), patients often notice a loose screw early, as a slight click when they bite or chew. Clinical signs are subtle at first: a gap or fluid at the margin and movement of the restoration in excursions.
A zirconia restoration bonded to a Ti-base can debond from the base while the screw stays tight. Pitta and colleagues (The International Journal of Prosthodontics, 2021) report that failure on a multi-unit prosthesis on Ti-bases after short service, and note that lab studies had already flagged the risk. The restoration spins or comes off; the base stays on the implant.
A radiograph sorts out the rest, as long as the beam is perpendicular to the connection. In an in vitro study by Sharkey and colleagues (The International Journal of Oral & Maxillofacial Implants, 2011), gaps of 12.7 µm were reliably seen with the beam up to 5° off perpendicular. At 15° off, only gaps of 51 µm or more were reliably detected.
What makes a screw loosen again?
A screw that loosens twice in the same place has a cause beyond torque. The table lists the causes described in the literature, the sign each one leaves and what to do.
| cause | sign | what to do | source |
|---|---|---|---|
| Under-torquing or no retightening | loosening with nothing else found: no gap on the radiograph, occlusion fine | retighten with a torque driver to the manufacturer’s value, then again after 10 minutes | Huang and Wang, 2019; Siamos et al., 2002 |
| Component not seated or bridge without passive fit | gap on a perpendicular radiograph; loosening soon after delivery, sometimes on several screws | fix the seating before retightening; on a full arch, test passive fit | Sharkey et al., 2011; Katsoulis et al., 2017 |
| Heavy or lateral occlusal contact | strong mark on the crown in light closure or in excursions | adjust to contact in firm closure only, no excursive contact on posterior crowns | Stilwell, 2024 |
| Cantilever | cantilevered restoration that keeps loosening after the occlusion is adjusted | review cantilever length and design with the lab | Huang and Wang, 2019 |
| Bruxism | wear facets, chipping, scalloped tongue borders | hard occlusal splint and a reviewed occlusal scheme | Vidal et al., 2025; Stilwell, 2024 |
| Screw tightened and loosened many times | worn threads or head, torque that won’t hold | replace the screw | Shui et al., 2026 |
| Restoration debonded from the Ti-base | restoration spins or comes off while the base stays on the implant | remove it and send restoration and base to the lab | Pitta et al., 2021 |
On a full-arch bridge, a screw that keeps loosening on one abutment is a reason to test passive fit. In Katsoulis and colleagues’ review (European Journal of Oral Implantology, 2017), misfit slightly raised the risk of screw-related complications, with no confirmed effect on bone. The guide on full-arch bridges that won’t seat shows how to test.
Retighten or replace the screw?
Retighten a screw that has loosened once if its threads and head are intact; replace it when it shows wear or has already been tightened and loosened many times. In Xiao and colleagues’ systematic review (The International Journal of Oral & Maxillofacial Implants, 2017), the literature was thin. Still, the 44 loose screws reported in two studies stayed tight after a single retightening.
The authors don’t recommend routine screw replacement. The limit shows up in lab testing. In Shui and colleagues’ review (Journal of Dentistry, 2026) of 16 in vitro studies, removal torque dropped after five tightening and loosening cycles. After ten it dropped further, and the threads showed wear.
Count every opening over the restoration’s life, not just the loosenings: try-ins, adjustments and repairs are cycles too. For a screw that has been through several, consider a new one.
When does a loose screw point to the occlusion?
A loose screw points to the occlusion when it comes back after correct retightening and a radiograph with no gap. In Hsu and colleagues’ review (The International Journal of Oral & Maxillofacial Implants, 2012), occlusal overload was considered the main cause of biomechanical complications, screw loosening among them. It was linked to bruxism.
Bruxism shows in the numbers. In Vidal and colleagues’ meta-analysis (Journal of Prosthodontics, 2025) of 13 studies, bruxism raised the odds of screw loosening 3.4-fold and of screw fracture 7.1-fold.
Before you retighten a third time, review the occlusal scheme. The guide on implant crown occlusal adjustment sets out the contact targets. For patients who brux, Stilwell’s review recommends a hard night-time appliance, such as a milled occlusal splint.
What should you tell the lab for a repair?
Talk to the lab before you ship and give the history: when the restoration went in, how many times the screw has loosened, the torque you used and what the radiograph showed. That way the case arrives with the right question.
Send the lab:
- The crown or bridge with its Ti-base and screw, each labeled.
- The radiograph taken perpendicular to the connection.
- Photos of the contacts marked in light closure, firm closure and excursions.
- Implant brand, system and platform, and the component model.
- If the restoration is being remade, a new scan with scan bodies, opposing arch and bite.
We work with Neodent and Titaniumfix components, both made in Brazil. If we made the restoration, give us the case in the portal (its screens are in Portuguese): the order holds the brand and platform you sent. If a full-arch acrylic bridge also has a broken or worn tooth, that goes in as a repair.
Ask how adjustments and remakes work before your first case. What you can prevent on the Rx, from scan bodies to the opposing arch, is in why implant crowns get remade. The restoration itself is on the zirconia implant crown page.
Related services
Open a portal accountFrequently asked questions
What torque should I use on an implant crown screw?
Whatever the component manufacturer's instructions say. It differs between abutment screws, Ti-base screws and multi-unit screws. Use a torque driver, not finger tightening. If you don't have the instructions for the component that came with the restoration, ask the lab for the brand and model used.
Can a cement-retained crown have a loose screw too?
Yes. On a cement-retained crown, the screw that loosens is the abutment screw underneath, and the whole unit gets play. Reaching it means cutting an access through the crown or removing it, which is one reason retention type matters at the planning stage.
What goes over the screw before I seal the access?
PTFE (Teflon) tape. It is inert and protects the screw head for the next re-entry, according to Stilwell's review in the British Dental Journal (2024). Composite goes on top, chosen for how soon you expect to reopen.
Sources
- Systematic review of the survival rate and the incidence of biological, technical, and aesthetic complications of single crowns on implants reported in longitudinal studies with a mean follow-up of 5 years · Clinical Oral Implants Research (Jung et al., 2012)
- Mechanism of and factors associated with the loosening of the implant abutment screw: a review · Journal of Esthetic and Restorative Dentistry (Huang and Wang, 2019)
- Relationship between implant preload and screw loosening on implant-supported prostheses · Journal of Oral Implantology (Siamos et al., 2002)
- A radiographic analysis of implant component misfit · The International Journal of Oral & Maxillofacial Implants (Sharkey et al., 2011)
- Opinions regarding reuse or replacement of implant prosthesis retaining screws: a systematic review · The International Journal of Oral & Maxillofacial Implants (Xiao et al., 2017)
- Impact of repeated tightening and loosening on the implant-prosthetic screw joint: a systematic review and meta-analysis · Journal of Dentistry (Shui et al., 2026)
- Relationship between bruxism and different types of mechanical complications in implant-supported prosthesis: a systematic review with meta-analysis · Journal of Prosthodontics (Vidal et al., 2025)
- Occlusal considerations in maintaining health of implants and their restorations · British Dental Journal (Stilwell, 2024)
- Biomechanical implant treatment complications: a systematic review of clinical studies of implants with at least 1 year of functional loading · The International Journal of Oral & Maxillofacial Implants (Hsu et al., 2012)
- Technical complication of a zirconia multiple-unit FDP supported by titanium base abutments: case report on a bonding failure and treatment alternative · The International Journal of Prosthodontics (Pitta et al., 2021)
- Misfit of implant prostheses and its impact on clinical outcomes. Definition, assessment and a systematic review of the literature · European Journal of Oral Implantology (Katsoulis et al., 2017)
One technical letter a month
Materials, protocols and what we see at the bench. No promotions, and you can leave anytime.